5 Sensor Installation
VIEW 370/373
18
(PC-000855-MA Issue 6, December 2021)
ENVEA
UK Ltd
Sensor
Route the input cable through the cable gland nearest to connector
K1
. Make the
connections to the
K1
terminals from top to bottom as outlined below.
Pins (top to bottom):
Wire colour
Connector name
SCR
Cable screen
SENSOR:
Power/comms
connector
B (Comms)
Blue
A (Comms)
Green
- PWR (0V)
Black
+ PWR (24V)
Brown
At the Network Spur
At the spur, make the connections as outlined below.
Pins (L
–
R):
Wire colour
Connector name
0V
Black
SPUR:
any connector
Comms B
Blue
Comms A
Green
24V
Brown
SCR
Cable screen
Before refitting the cover lid, the sensor unit must be configured in accordance
with the setup instructions provided in the following sections.
5.3.3
Set Address Switch (SW100)
Each sensor must have a unique address on the network. The address is set by
means of a DIP switch (SW100).
Only switch positions 1 through 6 are used, giving valid address settings from 1
to 64. The illustration below shows address 5. The address switch uses binary
coding to calculate the required address.
Example: to set the sensor to address number 5: set switch 1 (binary 1) ON and
switch 3 (binary 4) ON (4 + 1 = 5).
5.3.4
Set Sensor Comms Mode (SW101)
The sensor offers two communications (comms) modes: ASCII and RTU. The
Comms mode is set by means of a DIP switch (SW101).