42
dataPAC User’s Guide
Chapter 2 - The dataPAC
Band Overall Amp.
Enable this option to display a calculated amplitude between the high and low cursors using
a Root-Sum-Squared (RSS) calculation. Band RSS Amplitude is displayed only when Dual
Cursors is selected.
Averaging Information
Enable this option to display the number of averages and type of averaging used to obtain
the displayed data. The number of averages and type of averaging is controlled by the route
information loaded into your dataPAC.
Machine Speed
Enable this option to display the machine speed at the top of the display, The machine speed
is downloaded to the dataPAC along with other route information but can be changed in the
Review Data mode using the Set Speed softkey. If it is a triggered measurement, the speed
is measured and updated. Also, you may set the speed in Data Collection using the <F3>
softkey for non-triggered measurements.
Window
Enable this option to display the type of windowing function applied to the time domain
data when calculating the FFT. The Window setup is downloaded from the host computer
along with other route information.
Resolution
Enable this option to display the total number of FFT bins and the bin width in Hz or CPM.
The setting for this item is downloaded from the host computer along with other route
information.
Fmin./Fmax.
Enable this option to display the frequency value for the lowest and highest frequency bins
in the FFT data. The setting for this item is downloaded from the host computer along with
other route information.
Style - Line Graph
Enable this option to display the FFT as a line graph, rather than a histogram (bar chart).
Date/Time Collected
Enable this option to display the collection date and time for the FFT.
Delta Frequency
With Dual Cursors active and this option enabled, the dataPAC displays the frequency
spacing between the high and low cursors.
Grid
Enable this option to set the FFT Display so that a grid overlays a graph when the screen
opens. This sets the grid on by default—a softkey allows you to turn the grid on and off at
display time.
Orders
Enable this option to set the frequency axis to default to orders of magnitude of the machine
speed, rather than in frequency units (Hz or CPM) when the screen opens. You can alternate,
at display time, between Orders and Frequency units using a softkey.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com