168
dataPAC User’s Guide
Chapter 5 - Balancing Measurements
4.
Start a balancing session by pressing either <F2> or <F4>. To enter new balancing data,
press <F4> (New Run) from the Balancing Options window. To continue a saved
balancing session, press <F2> (Continue Loaded Run). The Balancing window opens.
5.
Select the correct transducer setting if necessary. Press <F2> (Transducer) to open the
Transducer Options window. See “Installing and Defining dataPAC Transducers” on
page 48 for more information.
6.
Press <F1> (Positions) to enter the number of positions. The number of positions refers
to the number of rotor components, such as blades on a fan. The Edit Positions window
opens.
Enter 0 for a rotor without vanes or blades, or enter the number of vanes or blades.
When your entry is complete, press <DONE>. To cancel, press <SHIFT><F5> (Abort).