Two Plane Balancing with a Strobe Light
dataPAC User’s Guide
179
2.
With the Edit Trial Weight - Plane 1 window open, enter the magnitude of the trial
weight you are about to attach to the rotor in the first plane. When your entry is
complete, press <DONE>. To cancel, press <SHIFT><F5> (Abort).
3.
The dataPAC displays the Weight Units window. Highlight a unit of measure on the
displayed list and press <SELECT>. The units selected for the trial weights become the
default units for the correction and trim weights. Correction and trim weight units can
be changed; the dataPAC will convert between units.
Note:
You must always enter trial weight values in the same units of measure for each plane.
4.
Enter the plane 1 trial weight location. Use the edit window to enter the location of the
trial weight on the rotor in the first plane.
If the machine has no vanes or blades, the trial weight location should be specified in
degrees (0 - 359). If you entered the number of vanes or blades earlier in the procedure,
the location is specified in positions (vanes or blades), from 1 to the total number of
vanes or blades.
In either case, the position is measured from the reference mark in the direction
specified in the Weight Placement filed on the Balance Options screen. When your
entry is complete, press <DONE>. To cancel, press <SHIFT><F5> (Abort).