Proline t-mass A 150 HART
Technical data
Hauser
99
HART load
Min. 250 Ω
Dynamic variables
The measured variables can be freely assigned to the dynamic variables.
Measured variables for PV (primary dynamic variable)
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
Measured variables for SV, TV, QV (secondary, tertiary and quaternary
dynamic variable)
• Mass flow
• Corrected volume flow
• FAD volume flow
• Temperature
• Totalizer
17.5 Power supply
Terminal assignment
Transmitter
Connection version 4-20 mA HART, pulse/frequency/switching output
4
1
2
3
1
2
23
22
24
25
26
27
A0017178
1
Supply voltage
2
Signal transmission: Pulse/frequency/switching output
3
Signal transmission: 4-20 mA HART
4
Ground terminal for cable shield
Supply voltage
Order code for
"Power supply"
Terminal numbers
1 (L+)
2 (L-)
Option
D
DC 24 V (18 to 30 V)
Signal transmission
Order code for
"Output"
Terminal numbers
Output 1
Output 2
26 (+)
27 (-)
24 (+)
25 (-)
Option
A
4-20 mA HART active
-
Option
B
4-20 mA HART active
Pulse/frequency/switch output
Option
K
-
Pulse/frequency/switch output
Содержание Proline t-mass A 150