Operation
Proline t-mass A 150 HART
78
Hauser
11.4 Adapting the measuring device to the process
conditions
The following are available for this purpose:
• Basic settings using the
Setup
• Advanced settings using the
Advanced setup
→ 53 menu
11.5 Performing a totalizer reset
In the
Operation
submenu, 2 parameters with various options for resetting the totalizers
are available:
• Control totalizer
• Preset value
• Resetting the totalizer
Navigation path
"Display/operat." menu → Operation
Function scope of the "Control totalizer" parameter
Options
Description
Reset + hold
The totaling process is stopped and the totalizer is reset to 0.
hold
The totaling process is stopped and the totalizer is set to the defined start value in
the
Preset
parameter.
Reset + totalize
The totalizer is reset to 0 and the totaling process is restarted.
totalize
The totalizer is set to the defined start value in the
Preset
parameter and the
totaling process is restarted.
Function scope of the "Reset totalizer" parameter
Options
Description
Reset + totalize
Resets the totalizer to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
"Operation" submenu
Operation
→
Control totalizer
Preset value
Resetting the totalizer
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection/
User entry
Factory setting
Control totalizer
Control totalizer
value.
• Totalize
• Reset + hold
• hold
• Reset + totalize
• totalize
Totalize
Preset value
Specify start value
for totalizer.
Floating-point
number with sign
0