Diagnostics and troubleshooting
Proline t-mass A 150 HART
86
Hauser
Options
Description
Logbook entry only
The device continues to measure. The diagnostic message is entered in the Event
logbook (events list) submenu only and is not displayed in alternation with the measured
value display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
12.5 Overview of diagnostic information
Diagnosti
c number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the sensor
004
Sensor
Change sensor
F
Alarm
*
082
Data storage
1. Change main electronic module.
2. Change sensor.
F
Alarm
*
083
Memory content
1. Restart device.
2. Restore data.
3. Change sensor.
F
Alarm
*
Diagnosti
c number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the electronics
270
Main electronic failure Change main electronic module.
F
Alarm
271
Main electronic failure 1. Restart device.
2. Change main electronic module.
F
Alarm
272
Main electronic failure 1. Restart device.
2. Contact service.
F
Alarm
*
273
Main electronic failure 1. Emergency operation via display.
2. Change main electronics.
F
Alarm
*
282
Data storage
1. Restart device.
2. Contact service.
F
Alarm
283
Memory content
1. Transfer data or reset device.
2. Contact service.
F
Alarm
*
311
Electronic failure
1. Transfer data or reset device.
2. Contact service.
F
Alarm
*
311
Electronic failure
Maintenance required!
1. Do not perform reset.
2. Contact service.
M
Warning
*
Diagnostic behavior can be changed: Section 12.4 "Adapting the diagnostic behavior"
Diagnosti
c number
Short text
Remedy information
Status
signal
from the
factory
Diagnostic behavior
from the factory
Diagnostics for the configuration