Installation
Proline t-mass A 150 HART
22
Hauser
Has the correct orientation for the sensor been selected → 15?
• According to sensor type
• According to medium properties
• According to medium temperature
• According to process pressure
Does the arrow on the sensor match the direction of flow of the medium through the piping
→ 15?
Have sufficient inlet and outlet runs been provided upstream and downstream of the measuring
point?
Correctly aligned in the direction of flow?
Is the device adequately protected from precipitation and direct sunlight?
Is the device protected against overheating?
Is the device protected against excessive vibrations?
Check gas property (e.g. purity, dryness, cleanness).
Are the measuring point identification and labeling correct (visual inspection)?