Commissioning
Proline t-mass A 150 HART
58
Hauser
Navigation path
"Setup" menu → Advanced setup → Current output
Structure of the submenu
Current output
→
Assign current output
Mass flow unit
Corrected volume flow unit
FAD volume flow unit
Temperature unit
Current span
4mA value
20mA value
Failure mode
Failure current
Parameter overview with brief description
Parameter
Description
Selection/
User entry
Factory setting
Assign current output
Use this function to assign
a measured variable or
process variable to the
current output
Mass flow
Corrected volume flow
FAD volume flow
Temperature
Mass flow
Mass flow unit
Select the unit for mass.
Metric:
Gram: g/s; g/min; g/h;
g/day
Kilogram: kg/s; kg/min;
kg/h; kg/day
Metric ton: t/s; t/min; t/h;
t/day
US:
Ounce: oz/s; oz/min; oz/h;
oz/day
Pound: lb/s; lb/min; lb/h;
lb/day
ton: LTon/s; LTon/min;
LTon/h; LTon/day
ton: STon/s; STon/min;
STon/h; STon/day
Arbitrary unit (see TEXT
mass unit function): _ _ _
_/s; _ _ _ _/min; _ _ _
_/h; _ _ _ _/day
Country-dependent:
• kg/h
• lb/h