Proline t-mass A 150 HART
Diagnostics and troubleshooting
Hauser
85
A0017300-EN
Calling up remedy information
1. Navigate to the "Diagnostics" menu.
In the "Actual diagnostics" parameter, the diagnostic code is shown with a short
text.
2. On the right in the display range, hover the cursor over the "Actual diagnostics"
parameter.
A tool tip with remedial measures for the diagnostic number appears.
12.4 Adapting the diagnostic information
12.4.1 Adapting the diagnostic behavior
Each diagnostic number is assigned a specific diagnostic behavior at the factory. The user
can change this assignment for specific diagnostic numbers via the
Diagnostics No. xxx
parameter.
Navigation path
"Expert" menu → System → Diagnostic handling → Diagnostic behavior → Assign behavior
of diagnostic no. xxx
0723-1
Diagnostic no. 044
/ ../Diagn. behavior
Diagnostic no. 801
Diagnostic no. 274
Warning
A0014048-EN
You can assign the following options to the diagnostic number as the diagnostic behavior:
Options
Description
Alarm
Measurement is interrupted. The signal outputs assume the defined alarm condition. A
diagnostic message is generated.
Warning
The device continues to measure. A diagnostic message is generated.