Proline t-mass A 150 HART
Technical data
Hauser
101
Permitted temperature range
• –40 °C (–40 °F)...≥ 80 °C (176 °F)
• Minimum requirement: cable temperature range ≥ ambient tempe20 K
Signal cable
Current output
For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the
plant.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Supply voltage cable
Standard installation cable is sufficient.
17.6 Performance characteristics
Reference operating
conditions
• Calibration systems traceable to national standards
• Accredited in accordance with ISO/IEC 17025
• Air controlled to 24 °C ± 0.5 °C (75.2 °F ± 0.9 °F) at atmospheric pressure
• Humidity controlled < 40 % RH
Maximum measured error
o.r. = of reading; o.f.s. = of full scale value
• The full scale value depends upon the nominal diameter of the measuring device
and the max. flow of the calibration rig.
• Full scale values of the specified measuring range.→ 95
0
10
[% kg/h]
±10
0
50
80
[%]
90 100
±5
±15
20 30 40
60 70
±20
110 120 130 140 150
[% lb/h]
H
G
K, L
A0016921
7
Maximum measured error (% mass flow) as % of measured value/full scale value. G, H, K, L: Order code
options for "Calibration flow", see the following table