Proline t-mass 65 PROFIBUS DP/PA
Installation
Hauser
17
Perforated plate flow conditioners (19 hole) for use with flanged sensor 65F
This is a special Hauser version designed especially for use with the t-mass F sensor
(sizes DN 25 to 100 / 1" to 4"). The mounting hole patterns and sizing are of a multi-variant
design which means that one plate will fit different flange pressure classes e.g. Cl. 150 and
Cl. 300.
The flow conditioner and gaskets are fitted between the pipe flange and the measuring
device
7. Use only standard bolts which match the flange bolt hole to guarantee that
the flow conditioner is centered correctly.
The alignment notch must also be pointing in the same plane as the transmitter. Incorrect
installation of the flow conditioner will have a small effect on the measurement accuracy.
A0005116
Fig. 7:
Flow conditioner mounting arrangement (example)
1 = perforated plate flow conditioner, 2 = aeal/gasket, 3 = alignment notch, 4 = alignment in the same plane as the transmitter
Note
• Order the t-mass F sensor and the flow conditioner together to ensure that they are cali-
brated together. Joint calibration guarantees optimum performance. Ordering the flow
conditioner separately and using it with the measuring device will further increase mea-
surement uncertainty.
• The use of conditioners from other suppliers will affect the flow profile and pressure drop
and will have an adverse effect on performance.
• Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by the
customer.