Commissioning
Proline t-mass 65 PROFIBUS DP/PA
54
Hauser
Setting or viewing the active gas group
Go to the function SELECT GROUP and simply select 1 or 2 and then exit using ESC (+/-
keys
simultaneously). No save function is necessary.
!
Note!
The Quick Setup Gas function is not available if an in-situ calibration function has been per-
formed on the measuring device as the in-situ calibration curve refers to the sensor power
at each recorded flow point. Therefore, the programmed gas settings become redundant.
Performing the Quick Setup
1.
GAS GROUP
Use the
O
or
S
key to select the required GAS GROUP and continue with
F
.
– Set the ANALYZER INPUT to ON if a gas compensation input is being used
– Select the NUMBER OF GASES in the group from 1 to 8.
– select the GAS TYPE from the choose list.
– enter the MOLE % for each GAS TYPE (only if NUMBER OF GASES is 2 and more).
– The error message CHECK VALUES appears if the total mixture % does not equal
100%. Go back and check the mixture settings.
2.
SAVE CHANGES?
– Select YES to save the settings in GAS GROUP 1 or 2 and activate the last gas group
selected. Press
F
to continue or
– Select CANCEL to save the entered settings in buffer memory but not activate them
for measurement. If this function is selected, then it will be necessary to come back to
this gas group and save it at a later stage.
– Select DISCARD to clear the last changes and return to CONFIGURE GROUP to make
new settings.
3.
ANOTHER GAS GROUP?
– Select YES to continue to the CONFIGURE GROUP function. Use the
O
or
S
key to
select the desired GAS GROUP and proceed as per the above instructions.
– Select NO to exit to the Quick Setup.
!
Note!
More detailed information on the GAS GROUP programming can be found in the separate
"Description of Device Functions" manual (BA00114D/06/… see chapter GAS).