Proline Prosonic Flow P 500
6
Hauser
Sensor
Prosonic Flow P
• Measurement of:
• Pure liquids or slightly contaminated liquids
• Chemicals
• Solvents
• Liquid hydrocarbons
• Acids
• Alkalis
• Nominal diameter range: DN 15 to 4000 (½ to 160")
• Materials:
• Sensor holder:
Stainless steel 1.4301 (304), 1.4404 (316L)
• Sensor housing:
Stainless steel 1.4301 (304), 1.4404 (316L)
• Strapping band/bracket:
Stainless steel 1.4301 (304), 1.4404 (316L)
• Sensor contact surface:
Chemically stable plastic
DN 15 to 65 (½ to 2½")
A0011484
DN 50 to 4000 (2 to 160")
A0013475
1
Example: 1 sensor set with 2 traverses
Accessories for mounting
The necessary installation distances must be determined for the sensors. Information about the
medium, the pipe material used and the exact pipe dimensions is required to determine these values.
The values for the sound velocity of the following media, pipe materials and liner materials are saved
in the transmitter:
Medium
Pipe material
Liner
• Water
• Seawater
• Distilled water
• Ammonia NH3
• Benzene
• Ethanol
• Glycol
• Kerosene
• Milk
• Methanol
• User-specific liquid
• Carbon steel
• Graphite cast iron
• Stainless steel
• 1.4301 (UNS S30400)
• 1.4401 (UNS S31600)
• 1.4550 (UNS S34700)
• Hastelloy C
• PVC
• PE
• LDPE
• HDPE
• GFR
• PVDF
• PA
• PP
• PTFE
• Pyrex glass
• Asbestos cement
• Copper
• Unknown pipe material
• None
• Cement
• Rubber
• Epoxy resin
• Unknown liner material
Sensor set selection and arrangement
If mounting horizontally, always mount the sensor set so that it is offset at angle of +30° to the
top of the measuring pipe to avoid incorrect measurements caused by empty space at the top of
the pipe.