Proline Prosonic Flow P 500
50
Hauser
6. Screw the mounting rail onto the sensor holders.
A0043385
43
Mounting the sensor holders and mounting rail
7. Attach the coupling pad with the adhesive side facing down on the sensors (→ 77).
Alternatively coat the contact surfaces with an even layer of coupling gel (approx.
1 mm (0.04 in)), going from the groove through the center to the opposite edge.
A0043382
44
Coating the contact surfaces of the sensor with coupling gel (if there is no coupling pad)
8. Insert the sensor into the sensor holder.
9. Fit the sensor cover on the sensor holder and turn until the sensor cover engages with a click
and the arrows (▴ / ▾ "close") are pointing towards one another.
10. Insert the sensor cable into the sensor until the end stop.
A0043386
45
Mounting the sensor and connecting the sensor cable
The sensors can now be connected to the transmitter via the sensor cables and the error
message can be checked in the sensor check function. This completes the mounting
procedure.
• The visible measuring pipe surface must be clean (free from flaking paint and/or rust) to
ensure good acoustic contact.
• If the sensor is removed from the measuring pipe, it must be cleaned and new coupling gel
applied (if there is no coupling pad).
• On rough measuring pipe surfaces, the gaps in the rough surface must be filled with
sufficient amounts of coupling gel if the use of the coupling pad does not suffice (installation
quality check).