Proline Prosonic Flow P 500
Hauser
45
M6
50 mm
(2 in)
1
2
3
1
13 mm
A0043375
33
Holder with welded bolts
1
Welding seam
2
Locking nut
3
Hole diameter max. 8.7 mm (0.34 in)
Sensor installation – small nominal diameters DN 15 to 65 (½ to 2½")
Requirements
• The installation distance is known → 39
• The sensor holder is pre-assembled
Material
The following material is required for mounting:
• Sensor incl. adapter cable
• Sensor cable for connecting to the transmitter
• Coupling medium (coupling pad or coupling gel) for an acoustic connection between the sensor
and pipe
Procedure:
1. Set the distance between the sensors to the value determined for the sensor distance. Press the
movable sensor down slightly to move it.
b
a
A0043376
34
Distance between sensors as per the installation distance → 39
a
Sensor distance (back of sensor must touch the surface)
b
Sensor contact surfaces
2. Stick the coupling pad under the sensor to the measuring pipe or coat the contact surfaces of
the sensor (b) with an even layer of coupling gel (approx. 0.5 to 1 mm (0.02 to 0.04 in)).