Proline Prosonic Flow P 500
24
Hauser
Load
Ex connection data
Safety-related values
Order code for
"Output; input 1"
Output type
Safety-related values
"Output; input 1"
26 (+)
27 (–)
Option
BA
Current output
4 to 20 mA HART
U
N
= 30 V
DC
U
M
= 250 V
AC
Option
MA
Modbus RS485
U
N
= 30 V
DC
U
M
= 250 V
AC
Order code for
"Output; input 2";
"Output; input 3"
Output type
Safety-related values
Output; input 2
Output; input 3
24 (+)
25 (–)
22 (+)
23 (–)
Option
B
Current output
4 to 20 mA
U
N
= 30 V
DC
U
M
= 250 V
AC
Option
D
User-configurable input/
output
U
N
= 30 V
DC
U
M
= 250 V
AC
Option
E
Pulse/frequency/switch
output
U
N
= 30 V
DC
U
M
= 250 V
AC
Option
F
Pulse output, phase-
shifted
U
N
= 30 V
DC
U
M
= 250 V
AC
Option
H
Relay output
U
N
= 30 V
DC
I
N
=100 mA
DC
/500 mA
AC
U
M
= 250 V
AC
Option
I
Current input 4 to 20 mA U
N
= 30 V
DC
U
M
= 250 V
AC
Option
J
Status input
U
N
= 30 V
DC
U
M
= 250 V
AC
Intrinsically safe values
Order code for
"Output; input 1"
Output type
Intrinsically safe values
"Output; input 1"
26 (+)
27 (–)
Option
CA
Current output
4 to 20 mA HART Ex i
passive
U
i
= 30 V
l
i
= 100 mA
P
i
= 1.25 W
L
i
= 0 µH
C
i
= 6 nF
Option
CC
Current output
4 to 20 mA HART Ex i
active
Ex ia
U
0
= 21.8 V
l
0
= 90 mA
P
0
= 491 mW
L
0
= 4.1 mH (IIC)/15 mH
(IIB)
C
0
= 160 nF (IIC)/
1
160 nF (IIB)
U
i
= 30 V
l
i
= 10 mA
P
i
= 0.3 W
L
i
= 5 µH
C
i
= 6 nF
1)
Only available for the Zone 1; Class I, Division 1 version
Содержание Proline Prosonic Flow P 500
Страница 78: ......
Страница 79: ......
Страница 80: ...www addresses endress com 71508895 71508895...