Proline Prosonic Flow P 500
46
Hauser
3. Fit the sensor housing on the sensor holder.
A0043377
35
Fitting the sensor housing
4. Lock the bracket in place to fix the sensor housing on the sensor holder.
A0043378
36
Fixing the sensor housing
5. Connect the sensor cable to the adapter cable.
This completes the mounting procedure. The sensors can now be connected to the
transmitter via the connecting cables.
• The visible measuring pipe surface must be clean (free from flaking paint and/or rust) to
ensure good acoustic contact.
• If necessary, the holder and sensor housing can be secured with a screw/nut or a lead seal
(not supplied).
• The bracket can only be released using an auxiliary tool (e.g. screwdriver).