Proline Prosonic Flow P 500
14
Hauser
Input
Measured variable
Direct measured variables
• Volume flow
• Flow velocity
• Sound velocity
Calculated measured variables
Mass flow
Measuring range
v = 0 to 15 m/s (0 to 50 ft/s)
Measuring range depending on the sensor version.
To calculate the measuring range, use the
Applicator
sizing tool → 75
Operable flow range
Over 150 : 1
Input signal
Output and input variants
External measured values
The measuring device provides optional interfaces that enable the transmission of externally
measured variables (temperature, density) to the measuring device:
• Analog inputs 4-20 mA
• Digital inputs (via HART input or Modbus)
Various temperature measuring devices can be ordered from Hauser: see "Accessories"
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Current input
The measured values are written from the automation system to the measuring device via the
Digital communication
The measured values can be written from the automation system to the measuring via:
Modbus RS485
Current input 0/4 to 20 mA
Current input
0/4 to 20 mA (active/passive)
Current span
• 4 to 20 mA (active)
• 0/4 to 20 mA (passive)
Resolution
1 µA
Voltage drop
Typically: 0.6 to 2 V for 3.6 to 22 mA (passive)
Maximum input voltage
≤ 30 V (passive)
Open-circuit voltage
≤ 28.8 V (active)
Possible input variables
• Temperature
• Density