Function group
DIAGNOSIS
This function group can be used to define
the error current assignments for each
individual error and to disable (lock) field operation.
Function/display
Matrix
VH
1)
Description
Setting
Default
Customer
Error
assignment
VH
82
This function is used to define whether or not an error
current of 22 mA is output for the selected error code on
the current output.
Note:
If set to „
yes
“ (effective), an error current is
output for an error set by the MyPro.
A diagnostic code with the setting „
no
“
(not effective) has no effect on the current
output.
Value range:
Yes = 1
no
= 0
depending on
code
Error current
delay
VH
83
Sets the delay for a diagnostic code for which the error
current assignment „
yes
“ (effective) has been set. If such
a diagnostic code is set by the MyPro, this error becomes
effective as an error current after the delay defined here.
Note:
This delay applies to all
diagnostic codes.
Value range:
0 ... 30 s
2 s
Unlock/Lock
VH
89
Unlock/locks field operation (see chapter 4.5).
Note:
Field operation can be locked and unlocked
with the hand-held HART
®
terminal, with the
Commuwin II user interface or in the field.
Locking via the keypad has precedence over
software locking.
Note:
0097 = instrument unlocked (any other
entry locks the instrument)
9999 = instrument locked in field with key
combination “+” and “F” (unlocking via
the HART
®
interface or via operating
level 2 is not possible)
Value range:
0000 ... 9998
0097
■
Operating level 2
Functions marked this way are not available in the Profibus instrument variant.
Functional description
mypro CLM 431 / CLD 431 conductive
34