5.7
Service and simulation
Function group
SERVICE/SIMULATION
Function/display
Matrix
VH
1)
Description
Setting
Default
Customer
Diagnostic code
VH
90
Displays the active diagnostic codes with the highest
priority levels (see chapters 4.6.2 and 7.2).
Value range:
E 1 ... E 150
–
Device address
VH 92
Entry of device address.
Note:
Only available for PROFIBUS-PA.
Value range:
0 ... 126
126
Software version
VH
93
Displays the software version of the instrument.
depends on
instrument
version
Hardware version
VH
94
Displays the hardware version of the instrument.
depends on
instrument
version
Factory settings
(set default)
VH
95
This function is used to selectively reset the data areas of
the instrument to the factory settings.
Value range:
no
= 0 = no reset
I nSt
=
1= instrument (data specific to instrument)
SEnS
= 2 = sensor (data specific to sensor)
uSEr
= 3 = user (combination of 1+2)
no
Output current
simulation
VH
98
This function is used to switch the output current
simulation on or off.
Note:
Reset back to „
0
“ (simulation off) after
simulation.
Value range:
off
= 0 = off
on
= 1 = on
0
Current value
VH
99
Entry of a current value (independent of the measurement)
to be output at the current output.
Value range:
4.00 ... 22.00 mA
10 mA
■
Operating level 2
Functions marked this way are not available in the Profius variant.
LM431CE5.CHP
mypro CLM 431 / CLD 431 conductive
Functional description
Hauser
35