Function group
MAIN PARAMETERS
Function/display
Matrix
VH
1)
Description
Setting
Default
Customer
Set
4 mA value
VH
05
Entry of conductivity or resistance value for a current
value of 4 mA (see chapter 4.6.3).
Note:
A specific minimum distance from the 20 mA
value must be observed.
Value range:
0 ... 60 mS/cm
0 ... 200 M
Ω⋅
cm
Minimum distance:
Meas. value betw. 0 ... 199,9
µ
S/cm:
20
µ
S/cm / 25 k
Ω⋅
cm
Meas. value betw. 200 ... 1999
µ
S/cm:
200
µ
S/cm / 0,25 M
Ω⋅
cm
Meas. value betw. 2 ... 19,99 mS/cm:
2 mS/cm / 2,5 M
Ω⋅
cm
Meas. value > 20 mS/cm:
20 mS/cm / 25 M
Ω⋅
cm
0.0
µ
S/cm
Set
20 mA value
VH
06
Entry of conductivity or resistance value for a current
value of 20 mA (see chapter 4.6.3).
Note:
A specific minimum distance from the 4 mA
value must be observed.
Value range:
0 ... 60 mS/cm
0 ... 200 M
Ω⋅
cm
Minimum distance:
Meas. value betw. 0 ... 199,9
µ
S/cm:
20
µ
S/cm / 25 k
Ω⋅
cm
Meas. value betw. 200 ... 1999
µ
S/cm:
200
µ
S/cm / 0,25 M
Ω⋅
cm
Meas. value betw. 2 ... 19,99 mS/cm:
2 mS/cm / 2,5 M
Ω⋅
cm
Meas. value > 20 mS/cm:
20 mS/cm / 25 M
Ω⋅
cm
20.00
µ
S/cm
Operating mode
conductivity /
resistance
VH
09
Sets the transmitter operating mode to conductivity or
resistance measurement.
Note:
Changing the operating modes resets the
current output values to default.
conductivity
or
resistance
■
Operating level 2
Functions marked this way are not available in the Profibus instrument variant.
Functional description
mypro CLM 431 / CLD 431 conductive
28
Hauser