tions referred to the matrix representation of
the hand-held terminal or Commuwin II.
5.1
Main parameters
Function group
MAIN PARAMETERS
Function/display
Matrix
VH
1)
Description
Setting
Default
Customer
Measured value
VH
00
Displays the currently measured conductivity or
resistance value.
–
Temperature
VH
01
Displays the currently measured temperature value (see
chapter 4.6.1).
Note:
This is only displayed if the temperature
measurement is enabled (see VH 17: type of
temperature compensation).
Value range:
–35.0 ... 250.0 °C
–
Operating state
VH
02
Output of current operating state, e.g. indicates when field
calibration is currently being performed.
Note:
This function is only intended for operation with
the Commuwin II user interface or the
hand-held HART
®
terminal.
Commuwin:
Meas., cal. active, parameter setting
–
Unit of main
parameter
VH
03
Selects the unit for the parameter measured.
Note:
• This is only relevant for the data transfer
via the interface. The unit on the 4-digit field
display changes automatically based on
the auto range function.
• Depending on operating mode cond. / M
Ω
Value range:
µ
S/cm, mS/cm, S/m
k
Ω⋅
cm, M
Ω⋅
cm
µ
S/cm
or
M
Ω⋅
cm
Input damping
VH
04
This function describes the transmitter’s response to the
input signal. The value entered here corresponds to the
number of samples used for averaging.
Value range:
1 ... 10
3
■
Operating level 2
LM431CE5.CHP
mypro CLM 431 / CLD 431 conductive
Functional description
Hauser
27