Function group
BASIC FUNCTIONS
Function/display
Matrix
VH
1)
Description
Setting
Default
Customer
Temperature
compensation
type
VH
17
Switches the temperature measurement on/off and
toggles between manual and automatic temperature
compensation (MTC/ATC).
Note:
• If set to „off+MTC“, the preset MTC
temperature is used for compensation
• If set to „on+MTC“, the temperature can be
additionally measured via a temperature
sensor and output via the HART
®
interface
or in the field
• If set to „on+ATC“, the value measured by
the temp. sensor is used for compensation
Value range:
oFF
= off + MTC
oFF
.
t
= on + MTC
on
.
T
= on + ATC
on + ATC
MTC temp.
VH
18
Entry of reference temperature for manual temperature
compensation.
Value range:
–35.0 ... 250.0 °C
25.0 °C
Temperature
offset
VH
19
Adjustment of temperature measurement by an offset
value. Entry of an offset value or of the offset determined
during calibration (at operating level 1, this value can only
be viewed but not edited).
Note:
Only exists if temperature measurement
has been switched on (see VH 17
„Temperature compensation type“).
Value range:
–20.0 ... 20.0 °C
0.0 °C
■
Operating level 2
Functional description
mypro CLM 431 / CLD 431 conductive
30
Hauser