Failure
no.
Display
Measures
Error current
assignment
(default)
E80
Current output parameter range too
small
Spread current output.
not active
E100
Current simulation active
Choose correct current output parameter.
not active
E101
Service function active
Switch off service function or switch instrument off and on
again.
not active
E106
Download active
Wait for download to be finished.
not active
E116
Download error
Repeat download.
active
E150
Distance between temperature values
in
α
value table too small or not
monotonously increasing
Enter correct values in
α
value table (minimum distance
between temperature values of 10 K required).
not active
LM431CE7.CHP
mypro CLM 431 / CLD 431 conductive
Troubleshooting
Hauser
43