4.6.2
Diagnostic parameter menu
The diagnostic parameters show the current
temperature offset (from temperature calibra-
tion) and the diagnostic codes (error mes-
sages) that are active, beginning with the
highest priority (Prio_1). If no other key is
pressed for 30 s, the instrument automatically
switches back to the measured value display.
4.6.3
Parameter settings
This function can be used to display and set
the values that are essential for start-up.
These are:
• Temperature coefficient (
α
value)
• Cell constant
• Meas. value for 4 mA current output
• Meas. value for 20 mA current output
The value being edited flashes. When the
desired value has been entered, it is
accepted with „
F
”, and the system advances
to the next parameter.
Adjustment ranges
Temperature coeff.
α
: 0.00 to 10.00 %/K
Cell constant C:
0.0025 to 99.99 cm
–1
Minimum distance
20
µ
S/cm / 200
µ
S/cm
4/20 mA value:
2 mS/cm / 20 mS/cm
(dep. on meas. range)
Note:
A minimum distance between the 4
and 20 mA values is required (see
chapter 10)
➞
an error is signalled if
this condition is not fulfilled.
Diagnostic codes (up to 5 messages)
Continue with „–“
Display of „E–––“ indicates ‘no error’
Temperature offset
For checking of and information on tem-
perature calibration (see chapter 4.6.4 )
ANZ3.CDR
Normal meas.
mode
Key
Display
Description
F
F
F
F
F
Temperature coefficient
Display of current
α
value
Edit with „
+
“ or „
–
“ (value flashes)
Cell constant
Display of current cell constant
Edit with „
+
“ or „
–
“ (value flashes).
Measured value at 4 mA
Display of current value at 4 mA
Edit with „
+
“ or „
–
“ (value flashes)
Measured value at 20 mA
Display of current value at 20 mA
Edit with „
+
“ or „
–
“ (value flashes)
ANZ4.CDR
Normal meas.
mode
Key
Display
Description
Operation
mypro CLM 431 / CLD 431 conductive
24
Hauser