Micropilot M FMR250 FOUNDATION Fieldbus
Trouble-shooting
Hauser
81
Error
Output
Possible cause
Remedy
If the surface is not
calm (e.g. filling,
emptying, agitator
running), the
measured value
jumps sporadically
to a higher level
L00-FMR2xxxx-19-00-00-en-015
L00-FMR2xxxx-19-00-00-en-016
Signal is weakened by
the rough surface – the
interference echoes are
sometimes stronger
1.
Carry out tank mapping
→
basic
setup
2.
Set the process cond. (004) to
"turb. surface" or "agitator"
3.
Increase the output damping
(058)
4.
Optimise the orientation
(
→
ä
84)
5.
If necessary, select a better
mounting position and/or larger
antenna (
→
ä
16)
During filling/
emptying the
measured value
jumps ownwards
L00-FMR2xxxx-19-00-00-en-017
Multiple echoes
yes
→
1.
Check the tank shape (002), e.g.
"dome ceiling" or "horizontal
cyl"
2.
In the range of the blocking dist.
(059) there is no echo evaluation
→
Adapt the value
3.
If possible, do not select central
installation position (
→
ä
16)
4.
Perhaps use a stilling well
E641 (loss of echo)
L00-FMR2xxxx-19-00-00-en-018
Level echo is too weak.
Possible causes:
• Rough surface due to
filling/ emptying
• Agitator running
• Foam
yes
→
1.
Check application arameters
(002), (003) and (004)
2.
Optimise alignment (
→
ä
84)
3.
If necessary, select a better
installation position and/or larger
antenna (
→
ä
16).
100%
0%
t
→
actual
expected
100%
0%
t
→
actual
expected
100%
0%
t
→
actual
expected
100%
E 641
0%
t
→
eingetreten
erwartet