Commissioning
Micropilot M FMR250 FOUNDATION Fieldbus
64
Hauser
6.6
Commissioning with a FOUNDATION Fieldbus
configuration tool
!
Note!
For commissioning of the device with a FOUNDATION Fieldbus configuration tool you need to
know the DEVICE_ID, which consists of the following parts:
whereby:
6.6.1
Fist setup
1.
Open the configuration tool and load the Device Descriptions (*.ffo, *.sym and - if required by
the tool - *.cff). Ensure you use the correct files (
→
ä
40).
2.
The first time it is connected, the device reports as follows:
FMR24x-FF-DevRev5-in-NIConfig
3.
Identify the device using the DEVICE_ID and assign the desired field device tag name
(PD_TAG).
Factory setting: PD_TAG = E+H_MICROPILOT_M_XXXXXXXX
Device_ID = 452B48100F-XXXXXXXX
452B48 ID code for Hauser
100F ID code for Micropilot M
XXXXXXXX Device serial number, as printed on the nameplate