Operation
Liquiline System CAT860
26
Hauser
7
Operation
8
Commissioning
8.1
Function check
WARNING
!
Risk of injury from medium leakage, incorrect supply voltage
Safety risks for staff and incorrect operation of the device
‣
Check all the connections to ensure that the device has been properly connected.
‣
Make sure that the supply voltage matches the voltage indicated on the nameplate.
8.2
Venting the diaphragm pump
First vent the diaphragm pump.
1. Open the vent valve.
2. Fill the pipe from the canister to the diaphragm pump completely with cleaning solution.
3. Close the vent valve.
a0022329
1
2
3
4
On-site controller
Sample pump forwards
Sample pump in reverse (hold down the operating key)
Backflush filter with air
Backflush filter with cleaner
Key functions
Press once
Function 1
= LED on
Press and hold key down
Function 2
= LED flashes
Press twice
Stop
= LED off
1
2
3
4
1
2
3
4