Technical data
Liquiline System CAT860
36
Hauser
14.3
Performance characteristics
14.3.1 Type of sampling
Control unit, Liquiline System CA80 analyzer
14.3.2 Filtrate quantity
• 5.5 to 16.5 ml/min
• Factory setting: 8.25 ml/min
All the values have been determined with new filters.
14.3.3 Suction height of peristaltic pump
Max. 5 m (16 ft)
14.3.4 Hose length, filter to pump
Max. 5 m (16 ft)
14.3.5 Hose length, pump to analyzer
Max. 30 m (98 ft)
14.4
Environment
14.4.1 Ambient temperature range
Unheated
• 5 to 50
˚
C (41 to 122
˚
F)
Heated
• -20 to 50
˚
C (-4 to 122
˚
F)
14.4.2 Storage temperature
-20 to 60
˚
C (-4 to 140
˚
F)
14.4.3 Humidity
10 to 95%, not condensating
14.4.4 Degree of protection
IP55
14.4.5 Pollution degree
The product is suitable for pollution degree 2.