Installation
Liquiline System CAT860
16
Hauser
Mount the filter as follows:
1. Screw the immersion tube connection (straight, 90
˚
) onto the immersion tube.
2. Where applicable, screw the quick release fastener onto the immersion tube connection
(optional).
3. Remove the thread adapter nut from the hose. The thread adapter nut is not required when
installing with an immersion tube.
4. Guide the "filter to pump" hose with the connection for sample preparation through the
splash protection cap from below.
5. Guide the "filter to pump" hose with the connection for the filter through the immersion tube
from above.
6. If you are using a quick release fastener, insert the inner sleeve into the quick release
fastener (see "Quick release fastener" section).
7. Connect the 4 mm, blue polytetrafluoro-ethylene sample hose to the filter (replacement
hose PTFE, 4 mm, black).
8. Screw the filter onto the immersion tube connection or onto the quick release fastener if one
is used.
Screw the pipes together finger-tight (without any gaps). The threads are lubricated and
provided with an O-ring.