Maintenance
Cleanfit CPA871
56
Hauser
Seal replacement in the rinse connection
4
5
4
5
A0030292
35
Replacing seals, Part 3
4
Lock nut
5
Rinse connection adapter
1. Release the lock nuts (pos. 4) using an open-ended wrench or socket wrench (AF 19
mm, in protective cover).
2. Unscrew the two rinse connection adapters (pos. 5) using an open-ended wrench or
socket wrench (AF 17 mm, in protective cover).
10
10
A0030315
36
Replacing seals, Part 3
10 O-rings, rinse connection adapter
3. Remove the O-rings indicated (pos. 10).
4. Apply a thin layer of grease (e.g., Klüber Paraliq GTE 703) to the new O-rings.
5. Insert the O-rings into the corresponding grooves.