![Endress+Hauser Cleanfit CPA871 Скачать руководство пользователя страница 83](http://html.mh-extra.com/html/endress-hauser/cleanfit-cpa871/cleanfit-cpa871_operating-instructions-manual_2407465083.webp)
Cleanfit CPA871
Technical data
Hauser
83
Manual drive, insertion/retraction up to 8 bar
p[bar]
T[°C]
20
80
60
40
T[°F]
p[psi]
0
32
0
0
16
232
100
120
140
68
104
140
176
212
248
284
8
116
A
p[pa]
0
1600000
800000
A0039156
52
Pressure temperature ratings for basic and immersion chamber version for materials 1.4404,
Alloy C22 and PEEK
A
Basic and immersion chamber version
p[bar]
T[°C]
20
80
60
40
T[°F]
p[psi]
0
32
0
0
16
232
100
120
140
68
104
140
176
212
248
284
8
116
4
58
A
B
p[pa]
0
1600000
800000
400000
A0039155
53
Pressure temperature ratings for basic version for materials PVDF and conductive PVDF
A
Immersion chamber version
B
Basic version