Maintenance
Cleanfit CPA871
58
Hauser
10
8
7
12
11
6
9
10
A0030293
38
Replacing all seals
6
7
8
9
10
11
12
Service chamber
Support housing
O-ring, support housing
O-rings, bottom of service chamber
O-rings, rinse connection adapter
O-rings, top of service chamber
O-ring, outer service chamber
2. Remove the support housing (pos. 7).
3. Pull the service chamber (pos. 6) out of the support housing.
4. Remove the O-rings indicated.
5. Apply a thin layer of grease (e.g., Klüber Paraliq GTE 703) to the new O-rings.
6. Insert the O-rings into the corresponding grooves.