Assembly connection PN16
Delivery
Rinse connections vary depending on the connection selected (G¼", NPT¼" or Swagelok)
Connections
Thread
Delivery
Dummy plug
M16, male thread
Mounted
Pressure retarder
M16, male thread to M16 female
thread
Mounted
Rinse connection outlet
G¼" or NPT 1/4", female thread, or
pipe connection
Mounted on pressure retarder
Rinse connection inlet
G¼" or NPT 1/4", female thread, or
pipe connection
Included
Changing the dummy plug/rinse connection inlet
If an (automatic) cleaning or calibration system is used, the dummy plug must be replaced
with the accompanying rinse connection inlet. The inlet and outlet must be fully connected.
1.
A0043258
Remove the dummy plug.
2. Replace the flat seal with the O-ring.
3.
A0047539
Screw in the rinse connection inlet.