![Endress+Hauser Cleanfit CPA871 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/endress-hauser/cleanfit-cpa871/cleanfit-cpa871_operating-instructions-manual_2407465034.webp)
Mounting
Cleanfit CPA871
34
Hauser
5.2.2
Sensor installation
Preparing the sensor and assembly
NOTICE
Danger of medium penetrating if a defective sensor is installed.
‣
Inspect the sensor and use a new, intact sensor if necessary.
1
A0030154
23
Sensor installation
1
Thrust collar with O-ring
1. Remove protective cap from sensor. Make sure that the O-ring and thrust collar
(→ 23, item 1) are provided.
2. To aid installation, immerse the sensor shaft in water.
3. Move the assembly to the service position.
Installing and removing sensors
L
WARNING
Risk resulting from temperature, pressure and chemical composition!
‣
Establish pressure compensation in the service chamber.
‣
Before removing, clean and rinse the sensor adequately in the rinse chamber.
‣
Check process seals. (There can be no leakage of medium from the rinse chamber in limit
position when rinsing is disabled)