Mounting
Cleanfit CPA871
26
Hauser
Rinse connections
The service chamber connections make it possible to rinse the chamber (including the sensor)
with water or cleaning solution. The pressure difference between the sealing water and
process must not exceed 6 bar (87 psi).
The sealing water pressure must not exceed 8 bar (116 psi) in manual mode and 16 bar (232
psi) in pneumatic mode.
Install a pressure-reducing valve upstream if there is the possibility that water pressures
can exceed the specified sealing water pressure (8 bar (116 psi) or 16 bar (232 psi)).
NOTICE
Pressure difference too high between process and wastewater system or if rinse
connections are not properly connected.
Damage to seals
‣
Close rinse connections.
‣
Pipe-fit rinse connections.
‣
Use sealing water function.
Assignment of rinse connections
In the standard and immersion chamber version, the inlet and outlet of the service chamber
are fixed. The outlet of the service chamber is located under the leakage hole. The leakage
hole is sealed with an M5 screw.
1
2
3
4
A0029621
19
Connection of service chamber in the immersion chamber version
1
Service chamber
2
Service chamber inlet, IN
3
Leakage hole
4
Service chamber outlet, OUT
Leakage hole, M5 thread, optional connection must be supplied by customer
Used for visual inspection.
If medium leaks out:
1. Switch off the process