Maintenance
Cleanfit CPA871
52
Hauser
2. Do not use chemical cleaning agents.
Such cleaning agents cause a potential to build up at the electrode which takes a
few hours to dissipate. The potential causes errors in the measurement.
3. Do not use abrasive cleaning agents.
These can cause irreparable damage to the sensor.
4. Perform another calibration following cleaning if necessary.
Clean the sensor:
• Before every calibration
• Regularly during operation
• Before returning it for repairs
You can remove the sensor and clean it manually, or you can clean it in automatic mode
using the rinse water connection.
In event of minor deposit build-up:
1. Place the sensor in warm water.
2. Clean the sensor with a mild dishwashing detergent.
1)
only if the assembly is fitted out accordingly