Cleanfit CPA871
Operation
Hauser
47
Use a pneumatic pilot valve (4/2-way or 5/2-way) to insert/retract the assembly.
‣
Connect the two inputs.
If only one input is connected (e.g. for test purposes), the piston is blocked as the
sensor guide moves before the limit position lock is disabled.
Inserting/retracting the assembly if the compressed air supply fails
1
2
A0030306
32
Failure of compressed air supply
1
Limit position lock for service position
2
Limit position lock for measuring position
L
CAUTION
Risk of injury due to high medium pressure
‣
Depressurize the system.
If the compressed air supply fails, you can still move the assembly manually. Proceed as
follows here:
1. Use an open-ended wrench (AF 17 mm) to unscrew both limit position locks (item 1
and 2).
2. Move the assembly to the desired position.
3. Screw the limit position lock back in.