Electrical connection
Hauser
23
[ ]
W
20
30
10
0
1022
587
152
U
[V]
1
2
R
R
L
L
max
max
£
-
U
6.5V
23mA
B
B
A0031107
1
Power supply 10 to 30 V DC
2
R
Lmax
maximum load resistance
U
B
Supply voltage
If load is too great:
• failure current is output and "S803" displayed (output: MIN alarm current)
• Periodic checking to establish if it is possible to quit fault state
• In order to guarantee sufficient terminal voltage, a maximum load resistance RL
(including line resistance) must not be exceeded depending on the supply voltage UB of
the supply unit.
6.4
Post-connection check
Is the device or cable undamaged (visual check)?
Do the cables comply with the requirements?
Do the cables have adequate strain relief?
Are all the cable glands installed, firmly tightened and leak-tight?
Does the supply voltage match the specifications on the nameplate?
Is the terminal assignment correct?
If required: has protective ground connection been established?
If supply voltage is present: is the device ready for operation and do values appear on the display module or
is the green status LED lit?