Description of device parameters
Ceraphant PTC31B, PTP31B, PTP33B IO-Link
64
Hauser
1
2
0
SP1
RP1
0
HNO
p
t
1
HNC
A0034025
0
0-signal. Output open in quiescent state.
1
1-signal. Output closed in quiescent state.
2
Hysteresis
SP1 Switch point
RP1 Switchback point
HNO NO contact
HNC NC contact
To prevent switch-on and switch-off if values are around the switch point "SP1" or
switchback point "RP1", a delay can be set for the relevant points. In this regard, see
the
Switching delay time, output 1 (dS1)
and
Switchback delay time, output 1
(dR1)
parameter descriptions.