Installation
Ceraphant PTC31B, PTP31B, PTP33B IO-Link
16
Hauser
5
Installation
5.1
Mounting dimensions
For dimensions, see the "Mechanical construction" section in the Technical Information.
5.2
Installation conditions
• Moisture must not penetrate the housing when mounting the device, establishing the
electrical connection and during operation.
• Do not clean or touch process isolating diaphragms with hard and/or pointed objects.
• Do not remove process isolating diaphragm protection until shortly before installation.
• Always tighten the cable entry firmly.
• Point the cable and connector downwards where possible to prevent moisture from
entering (e.g. rain or condensation water).
• Protect housing against impact.
• For devices with gauge pressure sensor, the following applies:
NOTICE
If a heated device is cooled in the course of a cleaning process (by cold water, for
example), a vacuum develops for a short time causing moisture to penetrate the
sensor via the pressure compensation element (1).
Device could be destroyed!
‣
In the event of this happening, mount the device in such a way that the pressure
compensation element (1) is pointing downwards at an angle or to the side, if possible.
1
1
1
1
A0022252
5.3
Influence of the installation position
Any orientation is possible. However, the orientation may cause a zero point shift i.e. the
measured value does not show zero when the vessel is empty or partially full.
C
A
B
A0024708
Содержание Ceraphant PTP31B