EMTEK, LLC - R100 USERS MANUAL / AUG2019.REV01
63
Appendix B: Suggested R100/RAS Sanitization Procedure
B.1 Materials
•
Gloves (Sterile or Clean)
•
Low Particulate Shedding Wipes (e.g., Wipe All, Gamma Wipes)
•
Primary Disinfectants (e.g., Quaternary Ammonium Compounds)
Important Note
: Phenolic disinfectants should not be used on the R100 LCD/Touchscreen overlay, as it will cause
the overlay to become very brittle in a short period of time.
•
Secondary Disinfectants (e.g., 70% 0.2µm filtered or sterilized alcohol, EtOH, or IPA)
•
PLUGS: Blower inlet plug, Ethernet port plug, USB port plug, optional inlet cover cover
B.2 Sanitization Procedure
English
DANGER:
TO MINIMIZE THE CHANCE OF ELECTRIC SHOCK disconnect the R100
AC/DC power supply cord, and power off the unit during sanitization
DANGER:
DO NOT SUBMERSE the R100 Air Sampler in any liquids!
Français
DANGER
: AFIN DE RÉDUIRE LES RISQUES DE CHOC ÉLECTRIQUE débrancher le R100
Cordon d'alimentation AC / DC, et l'appareil hors tension pendant la désinfection
DANGER:
NE PAS immerger le contrôleur de R100 Air Samplre dans tous les liquides!