EMTEK, LLC - R100 USERS MANUAL / AUG2019.REV01
53
A.2
Materials
•
EMTEK R100 Air Sampler, RAS, and power supply, Calibrated for operation at 28.3 LPM (1 CFM),
or 100 LPM with applicable inlet cover.
NOTE
: If remote exhaust tubing will be used with the R100, it is ideal that the unit is calibrated with
the exhaust tubing in place.
• Standard 90 mm TSA Agar Test Plates
•
30-35°C and/or 20-25°C Incubators (or other appropriate temperature).
•
Gloves (Sterile or Clean)
•
Low Particulate Shedding Wipes (e.g., Wipe All, Gamma Wipes or equivalent/better)
•
Disinfectants (e.g., Quaternary Ammonium Compounds, 70% 0.2µm Filtered or sterilized alcohol)
•
Lab Marker and/or Pre-Printed Labels
•
Adhesive Tape
A.3
Maintenance Inspection
To assure appropriate operation of the R100 sampling assembly, prior to each days’ use, sampling personnel should
inspect the unit for any obvious physical defect. This inspection shall include but not be limited to a visual
inspection of the sample inlet cover inlet holes to assure that they are free of occlusions to assure proper sample
flow. If any maintenance need be performed, contact appropriate service or repair personnel.
IMPORTANT SAFETY PRECAUTIONS!!
English
TO MINIMIZE THE CHANCE OF ELECTRICAL HAZARD, assure that the AC/DC power supply
cord is not plugged in during sanitization.
DO NOT REMOVE THE PANELS or COVERS of the R100 to attempt any repairs. Contact
EMTEK, LLC or other qualified service personnel if the unit malfunctions.
DO NOT SUBMERSE the R100 in any liquids!
USE SUPPLIED PROTECTIVE PLUGS for Ethernet, USB, and Blower inlet ports.
TAKE ALL OTHER STANDARD ELECTRICAL SAFETY PRECAUTIONS when operating the
R100 Air Sampler.
Français
AFIN DE RÉDUIRE LES RISQUES DE RISQUE ÉLECTRIQUE, s'assurer que le cordon
d'alimentation AC / DC n'est pas branché au cours de désinfection.