EMTEK, LLC - R100 USERS MANUAL / AUG2019.REV01
38
LINE 3:
ADMIN OPTIONS: Allows access to ADMIN CONTROL screen. Requires entry of ADMIN
ID and Password to access ADMIN CONTROL screen.
Administrative Control Screen Options
Line 1
PROGRAMS: Controls the ability of the USER to create or delete Saved Programs
(OFF=User, ON=Admin Only)
Line 2
CLEAR DATA: Controls the ability of the USER to clear stored data
(OFF=User, ON=Admin Only).
DATE STORAGE:
The R100 will store up to 500 samples in its internal memory. A
warning will occur at power up, which requires acknowledgement, when memory hits 450
samples. The CPU will delete oldest sample runs to add new ones once 500 samples has been
stored in the memory or will prevent new runs from being taken once 500 runs have been
stored if the CLEAR DATA function is ON, preventing data runs from being deleted.
Line 3
SITE DESC: Controls the ability of the USER to add or delete site descriptions
(OFF=User, ON=Admin Only)
Line 4
CALIB DUE ALM: Controls the ability of the USER to turn the calibration due alarm
ON or OFF (OFF=User, ON=Admin Only). If set to ON by ADMIN will set notification
ON on the CAL screen as well.
Line 5
FLOW ALM: Controls the ability of the USER to turn the calibration due alarm ON or
OFF (OFF=User, ON=Admin Only)
Line 6
SET DATE/TIME: Controls the ability of the USER to modify the current Date and/or
Time on the unit. This function needs to be turned OFF to modify either parameter, and if
even set to OFF, an ADMIN will be needed to toggle the selection from OFF to ON to
OFF again, to modify either parameter once they have logged out, or powered the unit
off, to be able to alter this key parameter.
NOTE
: Once the options are set as desired, the R100 must be power cycled (Turned OFF then ON) to log
out the ADMIN and lock these functions as set, otherwise they can still be modified (turned ON or
OFF) by the user.