EMTEK, LLC - R100 USERS MANUAL / AUG2019.REV01
34
7.1.10 CALIBRATION (CALIB) Screen View
Calibration Screen Options
Line 1:
CALIB DUE DATE: Shows Due Date for Calibration of both the 28.3 and 100 LPM flow rates.
Entered through Calibration Software entry and is stored on the unit. Not alterable by user or
administrator.
Line 2:
CALIB DUE ALARM: If ON, at unit power up, one of two calibration status screens may
appear, if within a 14-day window of the calibration due date of the unit. This includes the
following:
“Calibration will expire in ## days”
notice window (Fig 7.1.10.1) will appear if within a 14-day
window of the calibration due date. This will count down with each day towards the calibration
due date, each time it is powered up. It will actually show a negative day value (e.g., -5 days…),
if the unit is operated beyond its calibration due date. To continue to operate the unit, just touch
the screen.
NOTICE CAL DUE
screen (Fig. 7.1.10.2) will appear, when the calibration due date is met or
exceeded. At that point, only an ADMIN will be able to bypass the notice and allow use of the
unit. But, at that point the unit should likely be submitted for calibration by an appropriate service
group.
NOTE:
This alarm function is controlled by the unit Administrator. It can only be
turned OFF if ADMIN OPTION control for the CAL DUE ALM is OFF (requires ADMIN ID and
Password).
Fig. 7.10.1.1 Fig. 7.10.1.2