Remote Exhaust Kit
The R100 may be used for monitoring of locations where it is not desirable to have the HEPA filtered
exhaust of the unit at the test location (e.g., LAF Hood, Isolator, Filling Line, Top of Cart of Bench, etc.).
As such, it is possible to exhaust the R100 remotely using the R100 Remote Exhaust Kit
(R100.RMTEXH.KIT), depicted below. Which includes an exhaust port adapter, and 6 feet (1.8 m) of
exhaust tubing (which may both be autoclaved, or chemically sanitized, as required). Longer lengths of
tubing may be used, but it is to be noted that the longer the length the tubing. The exhaust adapter is an
open design, and maximizes unit run time possible when using this option.
For use:
1)
S
imply thread the exhaust adapter into the exhaust port on the R100. This should be finger tight only. Tools
shoot not be used, or needed, to attach, or detach the fitting. Note, the nut, retaining the tubing, can be threaded
off first, and the male/male fitting can be attached to the exhaust port, and then the exhaust tubing/nut assembly
can be threaded into the R100 adapter fitting. This may be found to be an easier procedure for attachment.
2)
With the exhaust fitting/tubing attached place the other end of the exhaust tubing at the desired exhaust location
(e.g., below cart level, outside of the ISO 5 zone (e.g., LAF Hood, Isolator, etc.)