Operating & Maintenance Instructions
57
A2.2
Comparing LED pulses with substandard meter pulses
This method may be used where the test equipment has the facility to calculate meter errors
based on the pulse output from a substandard meter. It will be necessary to set the pulse value of
the meter under test (shown on the meter nameplate) into the meter test equipment.
The duration of each test must be at least 60 seconds
The number of LED pulses should be greater than:-
Itest x V x Pf x N x K Itest = Test current
1000 x 60
V = System voltage
Pf = Power factor of test load
N = Number of system elements
K = LED pulse value, impulses/kWh
A3
Checking Meter Registration Accuracy from Register Advances
For these methods the advance of the meter register is used rather than the LED Test Indicator.
A3.1
Using the 'Dial Test' values on the meter display
Meters may be set to show cumulative kWh and kvarh registers
with increased resolution by using Power master Unit instructions
via the optical or RS232 port. Dial test displays can be programmed
to 0, 1, 2, 3 or 4 decimal places.
Import
The display resolution will revert to normal when a further appropriate instruction (Setting power
down count to zero) is received, or after a programmable number of power cycles.
What you will need
A suitable supply and load or a meter test bench.
Substandard meter with a kWh (kvarh) display.
For each measured quantity:
1. Connect the meter and substandard meter to the supply
2. Instruct the meter to show register to suitable dial test resolution
3. Record initial values of the meter and substandard meter registers
4. Apply a suitable load to cause a significant register advance
5. Switch off the load to stop the register advancing. Leave the supply connected
6. Record the final register readings, compute the advances and compare the meter advance
with the substandard advance
Содержание A1120
Страница 2: ......
Страница 3: ...A1120 40 Programmable Electronic Polyphase Meter Chapter 1 Introduction M181 001 1B 5 2009 ...
Страница 4: ...Minor Change Record Date Modification 7 5 2009 Format change ...
Страница 7: ...A1120 40 Programmable Polyphase Meter Chapter 2 Operating Maintenance Instructions M181 001 2G 5 2010 ...
Страница 8: ......
Страница 52: ...44 A1120 40 Electronic Polyphase Meter Elster Metering Limited M181 001 2G 5 2010 Figure 3 Typical Nameplate ...
Страница 53: ...Operating Maintenance Instructions 45 Figure 4 Load Curves ...
Страница 70: ......
Страница 71: ......
Страница 73: ......
Страница 91: ......
Страница 156: ......