PURELAB Ultra Mk2 Version 3 11/08
Page 41
Step 4 - Fit Both By-pass Blocks
1. WET ‘O’ rings on by-pass blocks.
2. SLIDE by-passes into unit.
3. POSITION both by-passes onto appropriate spigots.
4. ENSURE each by-pass is locked in place.
5. SWITCH
ON
power.
Note: The
PURELAB Ultra
will automatically sense that
the bypasses are fitted and default to the
sanitization cycle.
6. ACCEPT starting the sanitization cycle.
The
PURELAB Ultra
will recirculate for 10 minutes.
The unit will then proceed to alternatively rinse to drain
and recirculate for a total of 44 minutes. Upon
completion the alarm will sound for 4 seconds.
The sanitization cycle is now complete and the unit will
now display the next sanitization reminder alarm date.
Step 5 - Remove By-pass Blocks
1. PUSH and TILT by-pass blocks.
2. SLIDE out of unit.
3. PLACE on hangers located inside the door.
Step 6 - Replace Purification Packs
1. INSERT Primary Purification Pack into left-hand side
(lower position).
Allow the
PURELAB Ultra
to recognize the Purification
Pack and ensure that it has been replaced into the
correct position.
2. INSERT Polishing Purification Pack into right hand side
of the compartment (upper position).
Allow the
PURELAB Ultra
to recognize the Purification
Pack and ensure that it has been replaced into the
correct position.
3. ENSURE Purification Packs are locked into position.
4. CLOSE front door.
Step 7 - Return to Normal Operation
1. PRESS
PROCESS button twice.
The unit has now been sanitized but should be allowed
to rinse to a suitable quality. For particularly sensitive
applications it is recommended that this occur
overnight.
Note:
Once a sanitization cycle has commenced the
PURELAB Ultra
cannot be used to purify
water until the cycle has been totally
completed. If the electricity supply to the
PURELAB Ultra
is interrupted whilst the cycle
is in process, when it is restarted the unit will
resume sanitization at point of interruption.
After sanitization the
PURELAB Ultra
will
default to continuous recirculation for a
twelve-hour period.
Spigots
Fit By-pass Block
Slide out
Tilt
Push
Bypass block
Hangers
Remove By-pass Block