Page 21
Step 13 - Auto-Restart
This allows the selection of the AUTO/MANUAL restart option. If
auto restart is selected the unit will automatically restart after a
temporary loss of power to the unit. In manual mode the unit will
remain in standby mode.
1. PRESS
SCROLL button to highlight appropriate box.
2. PRESS
TICK
9
button.
Step 14 - Water Feed Type
To ensure maximum efficiency the process is tailored for feed
waters with different pretreatment.
RO
Ideally
<
30
µ
S/cm and supplied from either an RO
(permeate) loop or reservoir.
DI
Filtered
DI loop or reservoir with recirculation
maintaining a quality
>
1M
Ω
-cm.
SDI
Service Deionisation with 0.2 micron prefilter feed
directly into the unit via a pressure regulator.
1. PRESS
SCROLL button to highlight appropriate box.
2. PRESS
TICK
9
button.
Step 15 - Reservoir Level (only shown if RO feed is selected)
When operating the system from a reservoir it is recommended
that a low-level switch be connected to protect the system from
running dry.
1. PRESS
SCROLL button to highlight appropriate box.
2. PRESS
TICK
9
button.
The Controller Set-up is now complete.
Note:
To escape from any of the set-up screens
press the PROCESS button.
Auto-Restart Screen
Water Feed Type Screen
Reservoir Level Screen